Software: StepOne™ and StepOnePlus™ Software v2.3
See link to Readme file, StepOne™ Software v2.3 Release Notes.
A. General protocol for Sybr-Green-based qPCR
example, kit by QuantaBio qScript One-Step SYBR Green RT-qPCR, w/ ROX.
draft
Simplified StepOnePlus setup for a 96-well plate
StepOne Plus software, do set these options:
Experiment type: Comparative Ct (or Standard Curve if that is your design)
Chemistry: SYBR Green
Passive reference: ROX
Reaction volume: 20 µL
Plate type: 96-well
Per-well reaction mix, 20 µL
For each well:
10.0 µL 2X qScript One-Step SYBR Green Master Mix
0.4 µL qScript One-Step RT
0.4 µL Forward primer (10 µM)
0.4 µL Reverse primer (10 µM)
2.0 µL RNA template
6.8 µL nuclease-free water
Thermal cycling program
Program the run as:
1. 50°C for 10 min
Reverse transcription
2. 95°C for 5 min
Initial denaturation / enzyme activation
3. 40 cycles of:
95°C for 10 sec
60°C for 30 sec → collect fluorescence here
4. Melt curve / dissociation stage: ON
Plate setup suggestions
For a simple full plate:
samples in technical triplicate
include NTC wells for each primer pair
include minus-RT controls if checking for genomic DNA contamination
keep each primer pair grouped together on the plate
Practical reminder
Because this is a ROX kit, do not leave passive reference set to none or a low-ROX option. Set it explicitly to ROX.
Quick bench version
Per well:
10.0 master mix
0.4 RT
0.4 F primer
0.4 R primer
2.0 RNA
6.8 water
Program:
50°C 10 min
95°C 5 min
40 cycles:
95°C 10 sec
60°C 30 sec, read
melt curve on
B. Setting up High Resolution Melting Analysis (melt curve) on Step One Software
Start StepOnePlus software (shortcut on desktop), submit Username: [default ]
From the Menu, choose Advanced Set Up
-
Experiment Properties
-
Experiment Name, ex: HRM_PLATE001_
-
Save to appropriate folder on computer
-
-
Instrument: Stepone Plus (96 Wells)
-
Type of Experiment: Melt Curve
-
Reagents: SYBR
-
-
Plate Setup
-
Define Target ( Refers to Locus) Add/ Name target as needed
-
Define Samples (Sample refers to DNA, PCR products by number)
-
Assign targets and samples tab
-
Click on wells, click on checkbox by target (assign) and sample (assign) for each well used
-
-
-
RUN Method
-
Select reaction volume
-
Melt Curve Stage
-
ONLY CHANGE drop step 2 temp from 60° to 50° C
-
Step 1: 100% ramp up 95°, 15 seconds
-
Step 2: 100% ramp down 50°C, 1 minute
-
Step 3: to 3% °C ramp up 95°C, 15 seconds
-
-
-
Click Start Run
-
Click on Analyze button
-
At run end export file to XLS format, save to appropriate folder on computer
back to Instrument page